苏州聚友保利检测科技有限公司

National hotline:

0512-63230241

通用banner
Your current location 首 页 >> Album >> Calder Will

ECHOGRAPH 1090 Digital Ultrasonic Flaw Detector

2021-07-15
ECHOGRAPH 1090 Digital Ultrasonic Flaw Detector
详细介绍:

ECHOGRAPH 1090 Digital Ultrasonic Flaw Detector




Small, sturdy, digital-makes ultrasonic flaw detection easier


The new digital portable ultrasonic flaw detector ECHOGRAPH 1090 makes ultrasonic flaw detection easier.


This is due to:


 Lighter body


The large-area color LCD screen (143 x 79 mm) makes the A-scan display, measurement results and operating menus very eye-catching even from a distance.


Simple operation button design to avoid complicated menu operation


What is even more amazing is that due to the reasonable instrument settings and simple probe calibration methods, inexperienced personnel can quickly master the operation method.


The sturdy aluminum housing makes ECHOGRAPH 1090 have a thin thickness of 50mm and a light weight of 1.6 kg, including a lithium battery that can be used for 13 hours.


ECHOGRAPH 1090 has faster electronic processing components and a pulse repetition rate of up to 1500HZ. The two gates of the ECHOGRAPH 1090 can measure the echo sound path and amplitude separately. It has a matching independent indicator light. ECHOGRAPH 1090 has a larger A-scan memory.


Therefore, ECHOGRAPH 1090 is of course the most perfect handheld ultrasonic flaw detector. The beautiful and sturdy color-wrapped rubber protects the instrument from impact damage and rough or uneven surfaces.


The interface of ECHOGRAPH 1090 enables ECHOGRAPH 1090 to transfer data with the computer arbitrarily and apply it in a small automatic flaw detection system.


Outstanding features:


DGS/DAC version


DGS and DAC curves with 4 additional evaluation curves, directly displayed in


Amplitude evaluation relative to DGS/DAC (dBrel)


With 6.3" (16cm) A-scan TFT-LC display, the instrument can display clearly even in sunlight, 256 colors


Plain language and technical window design make operation extremely easy


5 programmable shortcut keys


Important parameters can be adjusted directly by shortcut keys


The sturdy aluminum chassis is wrapped with beautiful colored rubber so that the instrument can withstand harsh environments


Lithium battery that can be used for up to 13 hours has a safe fast charging function


Has a power saving mode


Different colors can be selected for measurement mode and menu operation


The operating menu language, parameters and help prompts can be re-stored in the instrument through the computer.


A variety of auxiliary settings, such as: adjust the screen display, probe delay, DAC and DGS settings


A-scan parameter reference


Dynamic echo curve


USB interface


8 Hz--1500 Hz pulse repetition rate


Instant output of gate measurement results


Trigger form: internal, external, one echo


Update and/or upgrade operating software via PC (CD-ROM, e-mail, download) ﹡


Manufactured and tested according to EN 12668-1


According to demand


Ordering information:


ECHOGRAPH 1090 Basic


Includes red rubber protective cover, rechargeable lithium battery, host/charger and instrument case 1090.101ECHOGRAPH 1090 DAC


Includes red rubber protective cover, rechargeable lithium battery, host/charger and instrument case 1090.201ECHOGRAPH 1090 DGS/DAC


Including red rubber protective cover, rechargeable lithium battery, host/charger and instrument case 1090.301 Accessories Carrying bag Including strap 6189.001 strap 6189.002 PC cable (USB)1657.702 software ECOM 90,


PC software for Windows XP/20001995.005 optional protective rubber sleeve (black) 6189.006 protective rubber sleeve (yellow) 6189.007 protective rubber sleeve (green) 6189.008




ECHOGRAPH 1090 technical data at a glance


Screen


Screen type

Color LC display emission to enable the instrument to display clearly in the sun

Screen size

143.4 x 79.3 mm

Resolution

400 x 240 pix, 256 from 262144 colours

A scan size

142 x 73.5 mm

Scale

Electronic generation, selectable on/off

Scale division

Coarse: horizontal tenths, vertical fifths

Fine: horizontal 50 divisions, vertical 25 divisions



A-scan image and digitization


Image refresh rate

50 Hz

A-scan image

Envelope

Reversible

Freeze

Dynamic echo curve

Zoom in through gate 1


RF description

Across the entire adjustable range

Rectify

Full wave, positive, negative, RF

Restrain

Manual selection: 0-99% full screen height, step 1% (linear)

Zoom in

Gate range (gate 1) to full screen width

A/D conversion

9 bits

Digitization process

Direct, w/ A/D converter

scanning frequency

80 MHz

Scan error during analog-to-digital conversion

<+/- 0.5% full screen height at 4 MHz

Delayed response

< 20 ms

Scan range


Time base range

2.5—4850mm steel

Speed of sound

100- 15000m/s 在1m/s Stepping

Delay time base scan

0-3000 mm at 0.1m

Time base linear

+/- 0.5% Screen width

Pulse repetition rate

8 Hz – 1500 Hz (depending on the time base range, from -85% to +50% in 1% steps)

Trigger mode

Internal, external, 1 second echo



launcher


Number of transmitters

2 (1 resolution, 1 power)

TX pulse waveform

Single phase (negative) narrowband pulse

TX impedance

10, 50, 220, w/o [Ohm]



Amplifier and attenuation


Number of frequency ranges

3 (LF- and RF range, broadband)

Adjustable attenuation

100 dB steps 0.1-1-2-6-12-20 dB



Echo evaluation, determined by the size of the defect


Echo height indicator

(Valid for all gates)


% Full screen height (FSH)

dBrel (DAV version)

dBrel (DAV version)

mm FBH (DGS/DAV version)


Echo sound path indicator

Acoustic path

Depth and projection distance and/or decrease

Depth and projection distance (with angle scan)

Resolution 0.1 mmSt




Distance amplitude adjustment (DAV version)


Points

max11

DAC- Curve movement

max 80 d B

Additional evaluation curve

max4 (to 15 d B DAC- curve)



DGS method (DGS/DAC version)

DGScurve

0—3mm FBH And bottom wave

Reference reflection

Bottom wave or FBH

Additional evaluation curve

max4 (to 15 d B DGS-curve)



Gate


Number of gates

2

Reaction time

Pulse repetition frequency (max1500 Hz)

Operation mode

Normal, upside down, off

Adjustment range

Gate start:

0-3000 mm in steps of 0.1 mm

Gate width:

0-1000 mm in steps of 0.1 mm



0 - 250 Occurrence rate

Output conversion go / no go

(All gates)



Level: TTL (5V), low level: ZA = 100 Ohm

Response accuracy: +/- 0.5% full screen height

Conversion lag: <0.5% full screen height

Switch output duration: 12ms


Optical (all gates)

Front panel 2 LED lights



input Output


USB interface

USB1 Interface to connect PC and printer

VGA output*

External monitor can be connected

Synchronous input/output

TTL-level (5V), low level:

Trigger threshold is about 2V

Trigger input/output is standard setting




multiple choice


measuring system

Optional mm, inch

Date and time

Embedded fixed clock

Language

German, English and other languages*, can be stored and modified via PC (text can be edited via PC)



Storage device


A-scan description

The current A-scan description can be stored via the "Freeze" button

Data settings

There are 224 data settings in the memory including A-scan, parameters, time and text* (8 lines with 37 letters per line)



power supply


Basic operation

Via the main unit 1808.501:

85-264 VAC, 47-63 Hz

Output: 10 VDC

Operating temperature: 0 degrees Celsius-+50 degrees Celsius

Storage temperature: -40 degrees Celsius-+85 degrees Celsius

Allowable humidity: 0-95%


Battery operation time (full charge)

Lithium battery powered

About 7h when the screen backlight is on

About 13h when the screen backlight is off


Power saving mode

Optional power saving mode ON/OFF

Automatic shut-down

When the power supply or battery is below the low line



Mechanical parameters


size (H x W x D)

166 x 201 x 50 mm Without protective shell

190 x 217 x 64 mm With protective shell

Front panel (H x W)

147 x 178 mm

weight

2.0 kg (including lithium battery and protective case)

1.6 kg (with lithium battery and without protective case)


interface

2 x Lemo 1 (For probe)

Output interface

PC: USB 1

D-type interface (9 pin), defect output and external trigger

Printer: USB mini B socket

Power interface: 5.5 mm AC socket

*According to requirements


Level: TTL (5V), low level: ZA = 100 Ohm

Response accuracy: +/- 0.5% full screen height

Conversion lag: <0.5% full screen height

Switch output duration: 12ms


标签